Features added in ES601 Version 3.0 (all language options):
• Enhanced leakage measurement protocols to satisfy requirements of the new third edition of the IEC60601 safety standard (IEC60601-1:2005)
• Added a test for defibrillator paddle leakage per requirements of IEC60601-2-4:2002
• Improved compliance of patient leakage testing under DIN VDE 0751-1:2001 and IEC62353:2007 standards
Feature added in Version 3.1 (all language options):
• With respect to ES601 Test Procedures and Tester AutoSequences, enhanced operation of 'C', 'R', and 'M' instructions for improved remote control of NIBP simulators such as Clinical Dynamics AccuPulse and Pronk SimCube. Refer to Datrend application note AN-22.
Feature added in Version 3.2 (English language option only):
• Added feature to ES601 System Settings menu for printing equipment "PM Due" labels via a Dymo SE300 adhesive label printer