Datrend vPad Rugged

Electrical Safety Analyser for Field Service Engineers

PRODUCT SUMMARY

  • Portable Electrical Safety Tester / Analyser
  • Designed for Field Service Engineers
  • Robust Crush Proof and Waterproof Case
  • Easy-to-read graphics and reports
  • Wireless Interface to Corporate CMMS System
  • Automated and Manual Electrical Safety Testing per IEC 62353, AAMI/ANSI ES-1, NFPA 99-2012 and IEC 60601

HOW TO ORDER

Please contact our sales department on 0845 345 4226, or via our contact form, to order or find out more about this product.

STANDARD VPAD APPS

Built on the Vision Pad Platform, and housed in an Extreme Duty case, the vPad-ES Rugged offers the functionality and scalability that you have come to expect from Vision Pad Technology products plus the added assurance of knowing your system will be protected in the harshest of environments.


Standard
  • Manual Electrical Safety Testing per IEC62353, AS 3551, AAMI/ANSI ES-1 & IEC 60601 standards
  • Built into a Crush Proof case that complies with ATA 300 Impact Test
  • Standard; waterproof to 3 feet for 30 minutes or more
  • Padlockable Powerclaw latching system with soft grip foldable handle
  • Hi-resolution Android tablet Graphical User Interface
  • USB connection to measurement base; Bluetooth connection to base optional
  • Bluetooth connectivity for transfer of files to and from local devices (PC, smartphone, other tablets, etc.)
  • Extensive internal storage
  • Unlimited user comments in test records
  • Integrated ECG & Arrhythmia Simulator
  • Internal test record storage
  • Prints test result reports to Wifi or Bluetooth printers (printer not included)
  • Increases quality, accuracy and consistency
Optional
  • Automated Electrical Safety Testing per IEC62353, AS 3551, AAMI/ANSI ES-1, IEC 60601 & NFPA99-2012 standards
  • Can be upgraded from a basic manual tester to full automation with CMMS connectivity by purchasing Apps
  • Built in Wireless enables transfer of data from remote sites to corporate CMMS systems
  • DACOM multi-device, serial control of external test instruments
  • Can be configured to retrieve test data from non-Datrend Devices
  • Complete test procedures available instantly
  • Automated Testing Sequences for defibrillators, infusion devices, patient monitors, and more
  • All Auto-sequences and Checklists are 100% editable for total procedural Customisation
  • Automatic "flagging" of test results that are outside of pre-established test limits
  • Prints inspection stickers to label printer (printer not included); supports specific models of Brady, Dymo, Zebra, ATM
  • Automation Reduces paperwork, eliminates handwritten records, standardises medical device testing
  • Automation Increases efficiency – may save in excess of 50% over manual testing & hand-written records
Display:
10.1" colour LCD (1280 x 800)
User Controls:
Capacitive touchscreen
Wired Connectivity:
Micro USB 2.0 Type B (shareable with base unit)
XBUS Port (RJ11-6) USB 2.0 Type A (host) - optional
Wireless Connectivity:
802.11 b/g/n
Bluetooth 2.1+ EDR
Modes of Operation:
Manual (standard)
Automatic (accessory option)
Power:
120 Volt power outlet: 90 to 132 V ac rms
47 to 63 Hz, 20 A max.
230 Volt power outlet: 180 to 264 V ac rms
47 to 63 Hz, 16 A max.
Weight:
7.0 lb (3.2 kg)
Size:
Base unit: 14.3 x 11.1 x 4.7 in.
(36 x 28 x 12 cm)
Tablet PC (detachable): 10.5 x 6.7 x 0.5 in.
(27 x 17 x 1.3 cm)
Test Standard Selections:
ANSI/AAMI ES-1
IEC60601-1
NFPA99-21012
IEC 62353
Voltage:
Mains voltage:
Range: 90 to 264 V rms
Accuracy: ±(2 % of reading + 0.2 V)
External (point-to-point) voltage:
Range: 0 to 300 V rms
Accuracy: ±(1 % FS + 0.2 V)
External (point-to-point) micropotential:
Ranges: 0 to 199.9 mV rms
200 to 1,999 mV rms
2,000 to 19,999 mV rms
Accuracy: ±(1 % of reading + 1 mV)
Ground Resistance:
Method: Four-terminal, fully isolated
Test Current: 1A pulsed, 0.2A rms
Range: 0.000 to 2.000 Ω
Accuracy: ±(1 % of reading + 0.02 Ω)
Equipment Current:
Ranges: 0 to 1.999 A ac rms
2.00 to 19.99 A ac rms
Accuracy ±(2 % reading + 0.2A)
Duty cycle 0 A to 10 A, continuous
10 A to 15 A, 7 min. on/3 min. off
15A to 20 A, 5 min. on/5 min. off
Equipment Power:
Range: 0 to 2400 watts
Accuracy: ±(5 % reading + 5W)
Chassis and Lead Leakage Current:
Modes: AC+DC (True-rms)
AC only
DC only
Patient Load Selection: AAMI ES1-1993
IEC 60601
Crest factor: <=3
Ranges: 0.0 to 199.9 μA
200 to 1,999 μA
2000 to 19,999 μA
Accuracy: DC to 1 kHz | ±(1 % of reading + 1μA)
1 to 100 kHz | ±(2 % of reading + 1μA)
100 kHz to 1 | MHz ±(5 % of reading + 1μA)
Isolation test voltage: 100 % ±5 % of AC supply
Calibration Test Points:
Resistance: 1 ± 0.01 Ω
Current: 100 ± 1 μA
ECG Performance Waveforms:
Output:
Amplitude 1 mV QRS into Lead II
Impedance 500 ohms
Accuracy:
Frequency ±1 %
Amplitude ±2%
Waveforms:
ECG Complex 30, 60, 120, 180, 240 and 300 BPM
Square wave 0.125 Hz, 2 Hz and 1 kHz
Pulse wave 63 msec
30PPM and 60PPM
Triangle wave 2 Hz
Sine wave 0.5, 10, 40, 50, 60 and 100 Hz
CMRR tests SQR 2Hz & 1KHz, PUL 4s, SIN 0.5, 50 & 60Hz
Arryhthmias VFIB, AFIB, SVT, VTACH, PVC, and ASYS
 
All specifications subject to change without notice.