Datrend vPad-IN

Next Generation Infant Incubator & Radiant Warmer Testing System

PRODUCT SUMMARY

  • The ONLY Incubator analyser to test to IEC 60601-2-19 and 2-21
  • Follows Edition 1 and Edition 2 of Standard
  • Built on the Vision Pad platform to give you greater flexibility and scalability
  • Wireless control of tests within incubator
  • Changes to test can be made without disturbing heat environment
  • Measures all parameters automatically following user defined protocols
  • Results can be merged with Safety test data and imported into commonly used CMMS system

HOW TO ORDER

Please contact our sales department on 0845 345 4226, or via our contact form, to order or find out more about this product.

STANDARD VPAD APPS

Introducing the vPad-IN, the next generation infant incubator and infant radiant warmer testing system that meets the requirements of the IEC Standards for incubators (IEC 60601-2-19) and radiant warmers (IEC 60601-2-21).

vPad-IN has all the functionality and scalability that you have come to expect from Vision Pad Technology products. Parameters are sensed inside the incubator environment and results are transmitted wirelessly to a 10" Android tablet for analysis and display. Test parameters and test protocols can be modified without disturbing the equilibrium of the heated environment, thereby minimising test times.


Compatible with closed, forced air convection incubators, and open, radiant infant warmers (optional)

  • Simultaneous measurement of humidity, airflow, sound, and 6 different temperatures
  • The only Tester available to test in compliance with IEC 60101-2-19 and IEC 60601-2-21
  • All sensors are independently movable, to be placed in positions required by the Standard
  • Sound level measurements may be taken outside the incubator for external alarm test
  • Results can be merged with vPad Safety Tests results and imported into commonly-used Equipment Management Systems
  • Optional precision controlled heater permits testing of body temperature sensor accuracy
Temperature:
Measurement Range: 0 deg C to 50 deg C
Accuracy: T1+/- 0.05 deg C
Resolution: 0.01 deg C
Convection: 5 “puck” sensors (T1-T5) - radiant warmer
Air Convection: 5 sensors (alternate “pos”’ T1 - T5) - incubator
1 Mattress Temperature Sensor 
Skin Sensor2: Sensor “oven” with thermal pad contact 
28° C to 40° C in 2° C increments 
Resolution: 0.01° C 
Accuracy: +/- 0.05° C 
Airflow:
Measurement Range: 0.1 to 1 m/s
Accuracy: +/- 0.1 m/s at 50% RH +/- 15%
Sound Level:
Measurement Range: 30 dbA to 90 dbA, Type II
Accuracy: +/- 5 dbA
Resolution: 0.1 dbA
Relative Humidity:
Measurement Range: 0 - 100% Non-condensing
Accuracy: +/- 3 % to 100% RH
Resolution: 0.1% RH
Data Retention:
Capacity: up to 48 hours per test
Sample rate: 1/min or 1/sec, automatic, parameter dependent 
Test Specifications:
Incubator Testing Specifications
  • • Measures temperature at 6 points as per IEC 60601-2-19 (sub-clause 50.102) 
  • • Measures humidity at the center of the enclosure per IEC 60601-2-19 (sub-clause 103.1) 
  • • Measures airflow at 4 points in the enclosure per IEC 60601-2-19 (sub-clause 104.1) 
  • • Measures sound level inside the incubator per IEC 60601-2-19 (sub-clause 102.1) 
  • • Measures sound level inside the incubator during alarm per IEC 60601-2-19 (sub-clause 102.2) 
  • • Measures sound level outside the incubator per IEC 60601-2-19 (sub-clause 102.3) 
  • • Measures incubator temperature at center point (A) of incubator as per IEC 60601-2-19 (subclause 50.102) 
  • • Measures average incubator temperature (AIT) per IEC 60601-2-19 (sub-clause 2.9.106) 
  • • Measures incubator temperature (IT) per IEC 60601-2-19 (sub-clause 2.9.105) 
  • • Measures average temperature (AT) per IEC 60601-2-19 (subclause 2.9.103) 
  • • Measures average temperature at 4 points per IEC 60601-2-19 (sub-clause 50.102) 
  • • Measures skin temperature sensor accuracy per IEC 60601-2-19 (sub-clause 50.104)2 
  • • Confirms skin temperature sensor control of incubator per IEC 60601-2-19 (sub-clause 50.105) 
  • • Confirms independent incubator temperature sensor accuracy per IEC 60601-2-19 (sub-clause 50.106) 
  • • Measures difference between average temperature and control temperature per IEC 60601-2-19 (sub-clause 50.107) 
  • • Measures actual warm-up time compared to operating manual per IEC 60601-2-19 (sub-clause 50.108) 
  • • Measures overshoot and recovery to steady state per IEC 60601-2-19 (sub-clause 50.109) 
  • • Confirms humidity reading per IEC 60601-2-19 (sub-clause 50.110)

Radiant Warmer Testing Specifications
  • • Measures skin temperature sensor accuracy per IEC 60601-2-21 (sub-clause 50.101)2 
  • • Measures average warmer temperature at 4 points per IEC 60601-2-21 (sub-clause 50.102) 
  • • Measures average warmer temperature at center point (A) of incubator as per IEC 60601-2-21(sub-clause 50.102) 
  • • Confirms skin temperature sensor control of warmer per IEC 60601-2-21 (sub-clause 50.103) 
  • • Confirms airflow is less than 0.1 m/s per IEC 60601-2-21 (subclause 50.102) 
  • • Measures sound level on the mattress per IEC 60601-2-21 (sub-clause 102.1) 
  • • Measures sound level on the mattress during alarm per IEC 60601-2-21 (sub-clause 102.1) 
  • • Measures sound level in front of the warmer per IEC 60601-2-21 (sub-clause 102.1) 
All specifications subject to change without notice.