Datrend Infutest 2000 Series E

Dual-Channel Infusion Device Analyser

PRODUCT SUMMARY

  • Test all types – dual-rate, pulsatile flow and PCA infusion pumps
  • Analyse up to four channels simultaneously
  • Unique “Synchro-Start” function – optimise IV Pump and Analyser accuracy

HOW TO ORDER

Please contact our sales department on 0845 345 4226, or via our contact form, to order or find out more about this product.

Posted by de Smit Medical Systems
1st March 2016

The best solution for infusion device inspection
Simultaneously inspect up to four single channel IV pumps or four independent channels of a multi-channel device in just seconds with our Infutest 2000 Series E and optional RSM. Better than 1% accuracy on all pumps, including Baxter Colleague and Alaris® Signature.

Conduct the full range of flow, pressure, volume and interval tests on practically any fluid delivery system including single rate, dual rate and patient controlled analgesia (PCA) devices. These tests are performed following easy-to-use manual or autosequence formats using the front panel keypad and graphical display.

Work smarter
Interface Infutest 2000E to our ES601Plus Automated Safety Analyser or other test systems such as a medTester 5000C with IVPUMP autosequence option.


End-of-infusion analysis (EIA)
Accurate volume-to-be-infused (VTBI) measurements are a snap. EIA calculates the volume infused at the instant the STOP key is pressed or when the autosequence preset test time has elapsed.

SynchroStart
Accurately captures the initial start-up flow dynamics of your IV pump, an invaluable feature when testing pumps with a pulsatile flow pattern. No need to attempt to start both the IV pump and the analyser at the same time to make this critical measurement. Infutest 2000 does it for you...automatically!

DTP-5 COMPANION SOFTWARE
Infutest 2000 comes complete with Windows (98/2000/XP) based data transfer program.
- (DTP-5) for quick and easy data transfer, graphical analysis, autosequence editing, plus organisation and printing of test results.

Single Rate Test (Continuous Flow Conditions)
Flow Measurement
Nominal range: 0.1 - 1200 ml/hr
Max. flow measurable and displayed: 1700 ml/hr
Min. flow measurable and displayed: 0.04 ml/hr
Max. display resolution: 0.001 ml/hr
Measuring time: 10 min. at 0.1 ml/hr / 20 sec. max. above 6 ml/hr
Accuracy: Average flow: +/- 1%, 0.1- 1200 ml/hr
Ranges: LOW - 0.1 to 170 ml/h
HIGH - 170 to 1200 ml/h
Range selection: Automatic
Internal nominal effective collection volume: LOW range: 0.014 ml
HIGH range: 1.1 ml
Volume Measurement
Range: 0 - 9999 ml, Max. display resolution: 0.001 ml
Accuracy: +/- 1%
Back Pressure Measurement
Range: 0 - 300 mmHg, Zero offset error:+/- 5 mmHg,
Display resolution: 1 mmHg, Accuracy:+/- 1% FS, +/- zero offset error
Elapsed Time Measurement
Range: 0 - 100 hours, Display format:hh:mm:ss, Accuracy: +0/-1second
 
Pulsatile Flow Conditions
Flow Measurement
Nominal range: 5 - 999 ml/hr
Max. flow measurable and displayed: 1700 ml/hr
Min. flow measurable and displayed: 2.75 ml/hr
Max. display resolution: 0.001 ml/hr
Measuring time: 14 min. at 5 ml/hr / 20 sec. max. above 200 ml/hr
Accuracy: Average flow +/- 1%
Range: HIGH range only (internally selected)
Internal nominal effective collection volume: 1.1 ml
Volume Measurement
Range: 0 - 9999 ml, Max. display resolution: 0.001 ml,
Accuracy: +/- 1%
 
Dual Rate Test (Continuous Flow Conditions ONLY)
Flow Measurement
Nominal range: 0.1 - 170 ml/hr
Max. flow measurable and displayed: 200 ml/hr
Min. flow measurable and displayed: 0.04 ml/hr
Max. display resolution: 0.001 ml/hr
Measuring time: 10 min. at 0.1 ml/hr / 20 sec. max.
above 6 ml/hr
Accuracy: Average flow: +/- 1%
Ranges: LOW range only
Delivery period determination: Automatic
Internal nominal effective collection volume: 0.014 ml
 
PCA Pump Test (Continuous Flow Conditions ONLY)
Flow Measurement, Bolus Delivery period
Nominal range: 0.1 - 170 ml/hr
Max. flow measurable and displayed: 200 ml/hr
Min. flow measurable and displayed: 0.04 ml/hr
Max. display resolution: 0.001 ml/hr
Measuring time: 10 min. at 0.1 ml/hr / 20 sec. max. above 6 ml/hr
Accuracy: Average flow: +/- 1%
Ranges: LOW range only
Lockout detection: Automatic
Internal nominal effective collection volume: 0.014 ml
Occlusion Pressure Test
Display Units: mmHg, psi / Display resolution:1 mmHg(0.1 psi)
Range: 0 - 2586 mmHg(0 - 50 psi)
Measuring time: 2 seconds
Zero offset error: +/- 5 mmHg (0.1 psi) Accuracy:+/- 1% +/- zero offset error
Lockout Time Measurement
Range: 0 - 100 minutes, Display format:mm:ss,
Accuracy: +/- 2 seconds
Nurse Call Detection
Nurse Call signal from pump sampled once per second.
Nurse Call signal can be produced by any of: relay contacts, open collector output, TTL output, RS-232 level, or 20 mA current loop.
Memory Type:
NVRAM (battery-backed)
Capacity:
2 channel: 900 flow and 700 pressure measurements per channel
4 channel: 450 flow and 350 pressure measurements per channel
Interface
User Interface: 40-character by 8-line backlit LCD, 4 front
panel soft-keys,
Internal beeper: LCD contrast control
Fluid interface:
Inputs A & B: Delrin twistlock, self-sealing / Common output A & B: Delrin twistlock
PCA Pump Trigger Outputs:
Mechanical One 1/4" female stereo phono jack per channel
Electrical Relay contacts rated at 120 VAC, 1 A; normally open
and normally closed contacts available.
Nurse Call Inputs:
Mechanical- One 1/4" female stereo phono jack per channel
Electrical- 50K-ohm differential input impedance.
Parallel Port
Centronix standard printer interface, DB-25F
Printer Driver
Epson MX/FX Series, 80 cps ,80 columns and 66 lines / page
Serial Port:
Mechanical - RS-232 / DB-25M
Electrical - RS-232C; bidirectional; CTS handshaking;
9600 baud, 8-None-1
Power Supply:
120 VAC 60 Hz @ 12 W (230/240 VAC 50 Hz
Europe/UK)
Environment:
15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II;
Dimensions:
12" W x 12" D x 6" H (30.5 cm W x 30.5 cm D x 15.2 cm H)
Weight:
10 lbs. (4.5 kg)
 
All specifications subject to change without notice.